Difference between Power Supply Current Measure IP and Power Supply Current Measure(IDDQ) IP.

Question

Please tell me the difference between Power Supply Current Measure IP and Power Supply Current Measure(IDDQ) IP.

Answer

Refer to the following table.
By combining Power Supply Current Measure IP and Multipoint Measurement Plugin, the same test as Power Supply Current Measure(IDDQ) IP can be performed.

Power Supply Current Measure IP Power Supply Current Measure(IDDQ) IP Power Supply Current Measure IP
+
Multipoint Measurement
Plugin
Measurement at SPA yes no yes
Measurement by loop section yes no yes
Measurement by JEC loop yes yes yes
Multipoint measurement no yes yes
Multiple time measurement in multipoint no no yes
I/O Control during measurement yes no yes
Acquisition of staticals
(average,maximum,minimum,3sigma)
no no yes
Individual specification for limits yes no yes
Visibility of measurement sequence yes no yes
PASS/FAIL judge of the pattern yes no yes
Support for External Power Supply Control Plugin yes no yes
Support for Arbitrary Waveform Applying Plugin yes no yes
Updated at
April 24, 2017